• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751NAS
Device Problems Improper or Incorrect Procedure or Method (2017); Calibration Problem (2890)
Patient Problem Hypoglycemia (1912)
Event Date 10/02/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
It was reported by the customer's parent that the customer received emergency medical assistance due to low blood glucose on (b)(6) 2017 with blood glucose of 28 mg/dl at the time of the incident.The customer was at 370 mg/dl at the time of the call.The customer was given a glucagon shot and intravenous fluids to treat.The customer was not wearing the insulin pump during the incident, within less than 48 hours.The customer stated that the pump gave them 60 units of insulin that was un-programmed.Troubleshooting was not completed as the customer declined.Customer was having sensor signal and calibration issues as well.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Findings: pump received with the operating currents within spec.And passed the self test, unexpected alarm error test, rewind, basic occlusion test, occlusion test, prime test, excessive no delivery test, displacement test, and the dat test.Pump was programmed with test minilink and the glucose sensor simulator.Pump communicated properly with glucose sensor simulator and displayed the programmed calibration blood glucose values, 100 mg/dl and 99 mg/dl properly on display graph.No weak signal alarms or sensor calibration errors noted.Pump was programmed with multiple boluses and monitored.All boluses delivered properly and were listed in the bolus history screen.Pump then was programmed with multiple basal profiles and monitored.All basal profiles delivered their indicated amounts and were verified in the daily total screen.The units left at pump display matched properly the units left at test reservoir.No delivery anomaly, bolus anomaly or basal anomaly noted during testing.Pump received with cracked reservoir tube lip, minor scratched lcd window, and scratched reservoir tube window.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6950238
MDR Text Key89341293
Report Number2032227-2017-58743
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503717
UDI-Public(01)00643169503717
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751NAS
Device Catalogue NumberMMT-751NAS
Device Lot NumberA4751NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2017
Date Manufacturer Received10/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age16 YR
Patient Weight126
-
-