It was reported that a patient¿s generator showed a low impedance message, upon interrogation.Clinic notes from a visit on (b)(6) 2017 were received for a referral that indicate the device has an issue and has since been turned off.The notes detail that the patient has been having an increase in ¿spells¿, clarified by the physician as partial seizures.Follow-up from the surgeon provided the low impedance was still present on the device.Additional relevant information has not been received to-date.No known surgery has occurred to-date.
|
Revision surgery occurred on (b)(6) 2017.During the surgery the physician found the lead to be severely twisted into a coiled knot.The generator pocket was then dissected and it was discovered that the generator¿s stitch was broken and the twisting of the lead had continued all the way down to the generator.The physician stated the patient had manually rolled the generator from the exterior, breaking the stitch and continued to manipulate and roll the generator within the pocket until the lead became severely twisted.The severe twisting caused the impedance issue, and when the surgeon removed and untwisted the lead, the impedance issue was corrected.The surgeon installed the same generator, repositioned the same lead, stitched two tie downs of the lead in the neck, re-stitched the generator, and stitched a tie down of the lead in the pocket.Three diagnostic tests returned with impedance within normal limits.
|