Model Number 25-40-46 |
Device Problem
No Flow (2991)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/20/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the s5 gas blender system.
The event occurred in (b)(6).
The device was sent to livanova (b)(4) for investigation.
The service engineer inspected the device and was able to confirm the reported issue.
During inspection the calibration showed a deviation in the o2 flow which was caused by a defective mass flow controller.
After the replacement of the defective part, the device was re-calibrated and subsequent tests were performed successfully.
A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
|
|
Manufacturer Narrative
|
There was no patient involvement.
Livanova (b)(4) manufactures the s5 gas blender system.
The event occurred in (b)(6).
This medwatch report is being filed on behalf of livanova (b)(4).
The device was returned to livanova (b)(6) where the issue was confirmed.
The unit has been returned to livanova (b)(4) for further investigation.
A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
If any additional information relevant to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that an s5 gas blender system had no oxygen flow.
There was no patient involvement.
|
|