Model Number N/A
Device Problem Difficult To Position (1467)
Patient Problem No Information (3190)
Event Date 09/15/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products- unknown locking pegs, cat#: ni lot#: ni.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a distal volar radius trauma plating procedure, two locking pegs would not lock into the distal holes on the wrist plate.The plate had to be removed and a new plate was utilized with the same locking pegs.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
(b)(4).This follow-up report is being filed to relay additional information which was unknown at the time of the initial medwatch.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination finds that there is damage to the threads of distal holes, likely from the attempts to lock the pins as reported.Functional or dimensional analysis could not be conducted, due to the damage to the threads.Device history record was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DVR ANATOMIC STANDARD HEAD RIGHT
Type of Device
PLATE, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6958032
MDR Text Key89680043
Report Number0001825034-2017-09099
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK111663
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberN/A
Device Catalogue NumberDVRAR
Device Lot NumberRM369G
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-