• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ALLURE RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3222
Device Problems Over-Sensing (1438); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/27/2017
Event Type  Injury  
Event Description
It was reported that during the check at the hospital, high pacing thresholds and oversensing were noted.The patient was asymptomatic.The device was electively explanted and replaced with medtronic to address pacing issue.The patient was stable.
 
Manufacturer Narrative
The reported field event of high capture threshold and oversensing was not confirmed in the laboratory.The device was tested on the bench and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6959792
MDR Text Key89672336
Report Number2938836-2017-33011
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model NumberPM3222
Device Lot NumberA000029623
Other Device ID Number05414734507905
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
-
-