The returned valve was patent.It met the requirements for reflux, pressure-flow, preimplantation, and leak testing.Therefore the conditions of the complaint could not be duplicated by laboratory personnel.The instructions for use that accompany the device instruct that when performing the patency check: a.Place the inlet connector of the valve into sterile isotonic saline.B.Depress the valve dome.C.Place a finger over the opening at the end of the outlet connector.D.Release the depressed dome.If fluid enters the reservoir, the inlet connector and flow control membrane valve are patent.* e.Remove finger from the outlet connector opening.F.Depress dome.If fluid flows out of the outlet connector, the valve is patent.* * note: it may be necessary to depress the valve dome more than once to initiate flow, especially if the ventricular catheter has been attached prior to patency testing.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|