Model Number H7493926024300
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2017
Event Type  Injury  
Manufacturer Narrative
Device is a combination product. (b)(4). Device evaluated by mfr : it is indicated that the device will not be returned for evaluation. If there is any further relevant information obtained, a supplemental medwatch will be filed. (b)(4).
 
Event Description
It was reported that stent migration occurred. The target lesion was located in the left anterior descending (lad) artery. After manually aspirating a visible thrombus in the lad, a 3. 00 x 24mm synergy ii drug eluting stent was deployed at 12 atmospheres. The balloon was fully deflated; however, upon attempting to remove the stent delivery system (sds), resistance was encountered and upon pulling sds again, it was noted that the stent moved proximally into the lad and left main segment. The stent was then post-dilated with a balloon catheter and was now well apposed to the lesion. The procedure was completed. No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Type of DeviceBIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6967336
MDR Text Key89914913
Report Number2134265-2017-10277
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08714729840237
UDI-Public08714729840237
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 10/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date11/16/2017
Device Model NumberH7493926024300
Device Catalogue Number39260-2430
Device Lot Number20005695
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/20/2017 Patient Sequence Number: 1
-
-