Model Number PM3222
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Information (3190)
Event Type  malfunction  
Event Description
It was reported that the device entered backup vvi mode.No further information is available at this time.
 
Manufacturer Narrative
The reported backup vvi (bvvi) was confirmed in the laboratory and was due to a power on reset (por).The device was tested on the bench and no anomalies were found.Power-on-reset could not be reproduced.The cause of the por could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key6969745
MDR Text Key90693845
Report Number2938836-2017-32870
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model NumberPM3222
Device Lot NumberA000042533
Other Device ID Number05414734507905
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-