The event log confirmed the reported all msd groups disabled alarm 36 minutes into the treatment.The investigation of the returned device on 26 september 2017 confirmed the msd shaft was fractured.The msd shaft was damaged with the rf common wire broken, which caused the msd failure.The optical microscopy evaluation noted a pinch mark on the msd shaft and damage 3cm from the end of the luer barb with exposing wires inside.It is likely that the complainant used a placement aid or tool during msd placement which contributed to the damage.Review of the device history record indicated that the device was manufactured according to specification and met all acceptance criteria prior to release.The instructions for use warn the user "do not deform or kink the msd during delivery into the iddc.If the msd is kinked at any time, do not attempt to use the msd, as kinking may lead to degraded performance or fracture during use." no serious injury was reported.The patient was reported as doing well.
|