Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and found that water was spilled into the device during the disinfection procedure, causing internal damage.The device was determined to be non-repairable and has been removed from service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|