Catalog Number 13200111 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/06/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
|
Event Description
|
Gamma 3 targeter missed on distal locking for a short nail procedure.Device was not damaged.Description: the patient was very obese, the distal locking hole missed, the drill hit the nail and would not advance smoothly through distal hole.The event was resolved by taking of the targeting device and performing a freehand locking technique in the distal most hole of the nail.
|
|
Manufacturer Narrative
|
The reported event that target device gamma3® was alleged of 'distal misdrilling / problems during drilling' could not be confirmed, since the device was not returned for evaluation and as it was mentioned that the material was not returned as the device was not damaged! no failure of the device as such was caused.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
|
Event Description
|
Gamma 3 targeter missed on distal locking for a short nail procedure.Device was not damaged.Description: the patient was very obese, the distal locking hole missed, the drill hit the nail and would not advance smoothly through distal hole.The event was resolved by taking of the targeting device and performing a freehand locking technique in the distal most hole of the nail.
|
|
Search Alerts/Recalls
|