Patient information was not provided.
Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.
The incident occurred in (b)(6).
This medwatch report is being filed on behalf of livanova (b)(4).
A livanova field service representative was dispatched to the facility to investigate.
The service representative was unable to reproduce the reported fault.
The drive unit of the centrifugal pump system was replaced as a precaution and subsequent functional testing was completed without issue.
The unit was returned to service.
If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
Evaluated on site by livanova technician.
|