Findings: all operating currents are within specification.No unexpected low battery or off no power alarms noted.Unit passed functional test including displacement test, rewind, basic occlusion, occlusion, prime/a33, and excessive no delivery alarm test.Unit functioned properly.Motor was tested outside the device and passed.Unit received with cracked case on display window corners, minor scratched lcd window, cracked reservoir tube lip.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.
|