|
Model Number M001182880 |
Device Problems
Break (1069); Device Damaged Prior to Use (2284)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/05/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the shaft was broken.A 135/10 renegade¿ hi-flo¿ was selected for use.During preparation, outside the patient's body, the carrier tube was hydrated before removing the catheter.However, it was noted that the shaft was broken upon removal from carrier tube.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.The tip, markerband, proximal shaft, and distal shaft were microscopically and tactilely inspected.Inspection revealed a separation in the outer shaft at 42.5cm from the strain relief, and the inner braiding stretched for 5 cm between the outer shaft separations.The outer shaft of the device was slightly tacky to the touch.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
|
|
Event Description
|
It was reported that the shaft was broken.A 135/10 renegade hi-flo was selected for use.During preparation, outside the patient's body, the carrier tube was hydrated before removing the catheter.However, it was noted that the shaft was broken upon removal from carrier tube.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Search Alerts/Recalls
|
|
|