• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S3 GAS BLENDER SYSTEM GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S3 GAS BLENDER SYSTEM GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-00
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 09/29/2017
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. The gas blender system 25-40-00 is not distributed in the usa, but it is similar to gas blender system 25-40-45, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the s3 gas blender system. The incident occurred in (b)(46). This medwatch report is being filed on behalf of livanova (b)(4). The device has been requested for return to livanova deutschland for investigation. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that an s3 gas blender system displayed an error message during priming. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameS3 GAS BLENDER SYSTEM
Type of DeviceGAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6976888
MDR Text Key90877139
Report Number9611109-2017-00864
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Other
Type of Report Initial
Report Date 10/25/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number25-40-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/20/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-