Catalog Number C-VH-1111 |
Device Problems
Device Operates Differently Than Expected (2913); Moisture or Humidity Problem (2986)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
|
|
Event Description
|
The hospital reported that during service/testing for an endoscopic vein harvesting procedure, bom 7mm extended length endoscope s/n (b)(4) was allowing moisture inside, the image was dark and cloudy.Hospital called seeking scope repair.(b)(4) contact was provided.
|
|
Manufacturer Narrative
|
(b)(4).This is a reusable oem device; therefore, a lot history review is not applicable.Based on the serial number provided, we were unable to find when the device was sold to the account.The certificate of conformance (c of c) was not available for review because, the reported serial number does not appear to be included in any of the batches received in (b)(4).
|
|
Event Description
|
The hospital reported that during service/testing for an endoscopic vein harvesting procedure, bom 7mm extended length endoscope s/n (b)(4) was allowing moisture inside, the image was dark and cloudy.Hospital called seeking scope repair.Schoelly imaging contact was provided.
|
|