COVIDIEN LP - SUPERDIMENSION INC SUPERDIMENSION INREACH SYSTEM; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number AAS00161-12
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/03/2017
Event Type  Injury  
Manufacturer Narrative
The device was not returned for evaluation.The site determined that the ct parameters were outside the recommended range.A review of the appropriate device history records of the superdimension console indicates this device was released meeting all quality specifications.Out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.If information is provided in the future, a supplemental report will be issued.
 
Event Description
During a superdimension enb procedure, the plan would not load on the system tower.The patient was under general anesthesia and the physician cancelled the case.
 
Manufacturer Narrative
Two planning files were returned and reviewed.Both files were able to load with planning application and procedure application on a procedure console.The event could not be reproduced.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERDIMENSION INREACH SYSTEM
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer (Section G)
COVIDIEN LP - SUPERDIMENSION INC
161 cheshire lane, suite 100
plymouth MN 55441
Manufacturer Contact
sharon murphy
161 cheshire lane
suite 100
minneapolis, MN 55441
2034925267
MDR Report Key6984208
MDR Text Key90451023
Report Number3004962788-2017-05107
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberAAS00161-12
Device Catalogue NumberAAS00161-12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/03/2017
Date Device Manufactured12/17/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-