MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE VIEW ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE VIEW ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1009-9212-000
Device Problem Detachment Of Device Component (1104)
Patient Problem No Patient Involvement (2645)
Event Date 10/04/2017
Event Type  Malfunction  
Manufacturer Narrative

A ge healthcare service representative performed a checkout of the equipment and confirmed the reported complaint. The manifold cover was replaced, and the unit was returned to service. No report of patient involvement. (b)(4).

 
Event Description

The hospital reported the unit's bag hose connection had become unglued and twisted off, resulting in an inability to manually ventilate. There was no report of patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAESPIRE VIEW
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 
MDR Report Key6991348
MDR Text Key91099842
Report Number2112667-2017-02018
Device Sequence Number1
Product Code BSZ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK092864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL,USER FACI
Reporter Occupation BIOMEDICAL ENGINEER
Type of Report Initial
Report Date 10/31/2017
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/31/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number1009-9212-000
OTHER Device ID NumberSEE BLOCK H10
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/04/2017
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured01/12/2017
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-