Model Number 24501879
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2017
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Some dark grey substance, e.G.Battery acid, was observed inside the battery compartment.Internally, there is some signs of foreign substance observed.Additional testing confirmed that batteries may leak when the battery is placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2017 to report an incident while pumping with her purely yours ultra breast pump that morning.She was in the middle of her pumping session when she heard a loud pop from the pump base and it suddenly ceased to function.Customer opened the battery compartment and noted dark fluid on the batteries and fluid pooling inside the compartment base.She denies coming in contact with the batteries or fluid therefore she was not burned or harmed.A replacement purely yours ultra pump base was overnight shipped to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS ULTRA
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key6991970
MDR Text Key91318970
Report Number3009974348-2017-00296
Device Sequence Number1
Product Code HGX
UDI-Device Identifier00810725021484
UDI-Public(01)00810725021484(21)
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 10/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24501879
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age30 YR
-
-