Catalog Number 8065751446
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
An ophthalmic surgeon reported that the trocar cannula could hardly be prick in the patient's eye during surgery.The product was not replaced and procedure completed with no patient harm.
 
Manufacturer Narrative
No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.One trocar cannula/hub assembly and one trocar handle/blade assembly were returned.The samples were visually inspected and was found to be non-conforming with damaged cutting edges.Penetration testing could not be performed due to the damage of the sample.The damage to the returned samples is consistent with damage that can occur when the blade contacts a hard surface such as the protective cap, improper handling, or contact with another instrument during surgery or set-up.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All trocar blades are 100% inspected.Any non-conformance's, such as damaged cutting edges, are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGEPLUS TROCAR/CANNULA SET
Type of Device
CANNULA, TROCAR, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6993986
MDR Text Key91474721
Report Number2028159-2017-04176
Device Sequence Number1
Product Code NGY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751446
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-