No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.One trocar cannula/hub assembly and one trocar handle/blade assembly were returned.The samples were visually inspected and was found to be non-conforming with damaged cutting edges.Penetration testing could not be performed due to the damage of the sample.The damage to the returned samples is consistent with damage that can occur when the blade contacts a hard surface such as the protective cap, improper handling, or contact with another instrument during surgery or set-up.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All trocar blades are 100% inspected.Any non-conformance's, such as damaged cutting edges, are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|