There was no lot number identified therefore a manufacturing device history record (dhr) review or product/process changes review for the involved lot number could not be performed.
However, all dhrs are reviewed for accuracy prior to product release.
One used single uvc catheter was received for testing and investigation.
The samples present residue from blood inside the tubing.
Additionally, the returned samples were inside a generic bag.
A visual inspection of the catheter revealed that the tubing was cut between 20-21 cm markings.
A magnified photograph was taken, and irregular cuts were visible on the tubing.
Leaks at strain relief junction can occur during normal use as a potential failure.
During the manufacturing a 100% leak proof testing is conducted as per procedure which would identify this issue in the catheter assembly.
Possible causes for the condition are: machine malfunction, unintentional customer misuse, product inspection inadequacy or not performed, or a sharp object was used that came into contact with the catheter are all possibilities.
Sharp objects should not be used with a catheter.
It is important to consider that the instructions for use warn: exercise caution when using sharp instruments near the catheter.
Do not use instruments with sharp or rough edges directly on the catheter since even a minor cut could tear or break the catheter.
Do not pinch or bend the catheter back to temporarily occlude the catheter.
This causes increased stress on the catheter which can lead to a leak or break.
Do not use clamps on umbilical vessel catheters] and continues, do not use alcohol, acetone, or alcohol containing antiseptics directly on the catheter.
Carefully check antiseptic solutions for alcohol or acetone.
These substances may cause irreversible damage to the polyurethane which can lead to a leak or break.
Ensure gloves or other surfaces which have alcohol on them are completely dry before touching or manipulating the catheter.
There are a number of alternatives on the field like exposure to chemical agents, proximity to heat sources or manipulation per se, that may lead to tubing tear.
Moreover, the condition found in this sample caused a leak which would be identified during assembly operations, since manufacturing performs 100% pressure testing during production.
Based on the available information this potential cause could not be discarded.
It must be noted that in-process controls such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed in the plant are in place to prevent nonconforming product from leaving the manufacturing operations.
This complaint will be used for tracking and trending purposes.
If information is provided in the future, a supplemental report will be issued.
|
The customer reports that during a procedure, a crack/ leak was noticed on the device between the 20-21 cm marking.
The catheter had to be removed from the patient.
In order to secure iv access, an unnecessary venipuncture had to be performed on the patient.
There was no injury to the patient as a result.
|