The returned anesthesia workstation has been simulated use tested in our laboratory environment.The reported issues with decreasing volumes have not been reproduced.The anesthesia workstation has been tested in different ventilation modes with different settings including settings made during the reported issue.The technical log does not have any recordings.The test log contains only successful system check outs (sco) including the sco performed on the morning prior to the reported event.The received logs have been evaluated both by our local application specialist as well as our in-house clinical experts.According to the trend log, the issues with a lower minute volume occurred on three occasions.During all of these three occasions, pressure control mode was used.The set pressure level above peep was 10cmh2o during all three occasions and the logs show further that the user never changed the pc above peep level even though mvi, tv, and cdyn gradually decreased over time during these three occasions.At the end of the three occasions, the operator switched ventilation mode to manual ventilation and since they are using afgo as manual ventilation mode.The experienced decrease in volumes, flow and cdyn (cdyn = dynamic compliance) are most likely related to the user not changing the pressure level above peep based on the actual condition of the patient at that time.Our conclusion is that there were no malfunctions of the anesthesia workstation during the reported time period.
|