Model Number H702496
Device Problem Computer Software Problem (1112)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/23/2017
Event Type  malfunction  
Event Description
During a comprehensive electrophysiology study with arrhythmia induction, the slave image froze.By changing screens the image would become live for thirty seconds and then freeze again.The case was continued by resetting the screen every time it froze.The case was completed however, there was some delay in the procedure time.There were no adverse consequences to the patient.
 
Manufacturer Narrative
Review of the case study logs did not reveal any software issues and there were no error messages correlating to the non-abbott remote monitor.Further information indicated that the site had adjusted the remote monitor configuration in attempts to work with non-abbott display hardware.Further communication with the customer revealed the issue was resolved the following day by the altering the resolution setting on the remote monitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE¿ PRECISION¿ SOFTWARE
Type of Device
ELECTROPHYSIOLOGY CARDIAC MAPPING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7004232
MDR Text Key91516074
Report Number2184149-2017-00048
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160210
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH702496
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age68 YR
Patient Weight97
-
-