(b)(4).
Upon completion of the investigation it was noted that the sample was received and reviewed.
No sign of burn, mark, or stain.
The reported " a surgical pattie was found to have a discolored stain" could not be verified.
However; samples were observed to contain 11 patties instead of 10.
The 11th pattie was missing string.
The dhr/manufacturing lot traveler was pulled and reviewed for lot h27272.
All information on the traveler indicates that the product was produced within specifications.
There were no engineering change orders or anything unusual relative to the manufacturing documentation of this lot.
Therefore, root cause could not be determined.
No corrective action will be taken at this time.
Based on the results of this investigation no further action is required.
Trends will be monitored for this and similar complaints.
At the present time this complaint is closed.
|