ABBOTT DIAGNOSTICS DIVISION CELL-DYN RUBY ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Model Number 08H67-01
Device Problem Pressure Problem (3012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/13/2017
Event Type  malfunction  
Manufacturer Narrative
A product correction letter was sent to cell-dyn ruby customers who have instruments with the affected printed circuit board assemblies (pcbas).A mandatory technical service bulletin (tsb) was issued on all impacted cell-dyn ruby analyzers.The mandatory tsb instructs field service to replace the pcba pump relay board.
 
Event Description
The distributor replaced the pcba pump relay board on the cell-dyn ruby analyzer per the technical service bulletin.There was no impact to patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN RUBY ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key7006497
MDR Text Key92211778
Report Number2919069-2017-00152
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740017170
UDI-Public(01)00380740017170
Combination Product (y/n)N
Reporter Country CodeHU
PMA/PMN Number
K061667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number08H67-01
Device Catalogue Number08H67-01
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/13/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Removal/Correction Number2919069-02/13/17-001-C
Patient Sequence Number1
-
-