LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Failure to Pump (1502)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/27/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). This medwatch report is being filed on behalf of livanova (b)(4). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and identified the patient pump to be faulty. The pump was replaced and subsequent functional verification testing was completed without further issues. The unit was returned to service. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report. Evaluated on site by livanova rep.
 
Event Description
Livanova (b)(4) received a report that the patient pump of a heater-cooler system 3t did not function during a procedure. There was no flow through the pump. There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7007560
MDR Text Key92200320
Report Number9611109-2017-00901
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Other
Type of Report Initial
Report Date 11/07/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number16-02-85
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-