Model Number V60
Device Problems Calibration Error (1078); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Event Description
Customer reported that the unit has error code messages of air flow sensor calibration data error and oxygen flow sensor calibration data error in the log.The customer reported there was no patient involvement.
 
Manufacturer Narrative
The product support troubleshot this issue with the customer over the phone.The customer was advised to run the performance verifications test (pvt) and to either replace the flow sensor or data acquisition (da) printed circuit board (pcb).Attempts have been made to try to obtain the repair information but no response received from the customer.No parts returned to failure investigation; therefore, the root cause of the reported issue could not be determined.If parts are returned, the complaint will be reopened and an investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key7008134
MDR Text Key91789028
Report Number2031642-2017-03486
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-