• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K2; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K2; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190610
Device Problem Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2017
Event Type  malfunction  
Manufacturer Narrative
A follow up will be submitted following evaluation.
 
Event Description
A dialysis biomedical technician reported that during treatment all treatment settings reset by themselves.There were no diagnostic messages or audible alarms.The machine power did not reset, only the treatment settings.He reported on debug screen 8 between 6:00 and 8:40 the ultra filtration volume (ufv) did not reset.It counted up to 300.8:40 was the last ufv entry.He was not able to find out what time the problem occurred.During follow up a facility biomed reported the patient finished treatment without issue on another machine and did not need any medical intervention and no blood loss.He did not have patient details.He reported that they tested the ufv with a simulated patient run and the issue recurred so they replaced the function board.That resolved the issue.The board was not retained.
 
Manufacturer Narrative
The alleged event is not confirmed.The device was not returned to the manufacturing plant for investigation.An investigation of the product history records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the product history review confirmed the labeling, material, and process controls were within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key7008381
MDR Text Key92316302
Report Number2937457-2017-01153
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100859
UDI-Public00840861100859
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Device AgeMO
Date Manufacturer Received11/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/03/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-