Catalog Number 9735669
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/12/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by manufacturer for evaluation.
 
Event Description
A site representative reported that, while in a functional endoscopic sinus surgery (fess), the monitor of the navigation system lost functionality after being manually manipulated by the representative.It was reported that the functionality was restored after thirty seconds and the site was able to complete the procedure.The reported issue occurred after instrument verification and registration were successfully completed.There was no reported delay to the procedure due to this issue.There was no impact on patient outcome.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 EM ENT SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7015652
MDR Text Key92639287
Report Number1723170-2017-04587
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169838918
UDI-Public00643169838918
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735669
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age13 YR
Patient Weight82
-
-