MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616SR
Device Problems Overheating of Device (1437); Computer Operating System Problem (2898); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/08/2017
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation. Product and no physical damage was observed. There was a relationship found between the returned device and the reported incident. Mdu failed functional tests with motor stall error and overheating when power cord was bent in the middle. Power cord assembly grew warm during testing. After troubleshooting, the cause of motor stall and overheating was observed to be a defective power cord assembly. A visual inspection was performed on the power cords external covering and no physical damage was observed, only a small kink near the strain relief. It was determined that the power cord has a shorted or open internal wire. Motor and hall board were tested and passed functional testing. The complaint investigation has concluded that this unit has succumbed to physical damage to power cord. Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported that the device was not working properly and the cord was very hot. No significant delay or patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVCE REPL MDU CNTRL PWRMX ELITE
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7031422
MDR Text Key141512964
Report Number1643264-2017-01726
Device Sequence Number1
Product Code HAB
UDI-Device Identifier00885556582305
UDI-Public(01)00885556582305
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/06/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No