MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616SR
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation. A visual inspection was performed on the product and no issue was observed. There was a relationship found between the returned device and the reported incident. Product failed functional testing with motor stall error and overheating. Cause of overheating and errors is a corroded motor/gearbox. The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion. The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly. A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing. Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported that the device gets hot to touch. No significant delay or patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVCE REPL MDU CNTRL PWRMX ELITE
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7040218
MDR Text Key122201496
Report Number1643264-2017-01736
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/09/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/08/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-