• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616S
Device Problem Device Displays Incorrect Message (2591)
Patient Problem Partial thickness (Second Degree) Burn (2694)
Event Date 10/23/2017
Event Type  Injury  
Manufacturer Narrative
Foreign zip code: (b)(6).
 
Event Description
It was reported that the mdu began to emit warning sound, showing short circuit message.The surgeon noticed that the device reached high tempetures.At the time of removing the field, 2 burns type second degree on the skin of the patient was noted.Fifteen minutes delay.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident of overheating.Complaint of a short circuit was not confirmed.Product failed functional testing with motor stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The motor/gearbox assembly could not be removed from the housing for further assessment due to corrosion.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVC REPL,MDU, HAND CNTRL, PWRMX
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7041594
MDR Text Key92357347
Report Number1643264-2017-01739
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010565105
UDI-Public(01)03596010565105
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616S
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age50 YR
-
-