Model Number 010807
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/27/2017
Event Type  malfunction  
Event Description
Medtronic received a report that the infant/ neonatal vitaline spring detached immediately after placing on the patient.There was no harm to the patient.
 
Manufacturer Narrative
Device evaluation: the sample was received for analysis and the reported event was confirmed.Visual and microscope inspection were performed and it was observed that the deformations were on the slider¿s lockers.The dhr couldn¿t been reviewed as the lot number of the complaint sample wasn¿t reported.The most probable root cause is that the deformations may be caused due to multiple use or due to misuse.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIDION MEDICAL
Type of Device
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ORIDION MEDICAL
7 hamarpeh st.
jerusalem 91235
MDR Report Key7042447
MDR Text Key93318625
Report Number8044004-2017-05010
Device Sequence Number1
Product Code CCK
Combination Product (y/n)N
PMA/PMN Number
K980327
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 04/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number010807
Device Catalogue Number010807
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Date Manufacturer Received03/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-