SMITH & NEPHEW, INC. RFB LIGHT SOURCE 500XL XENON IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568F
Device Problems Thermal Decomposition of Device (1071); Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/03/2017
Event Type  malfunction  
Manufacturer Narrative
There was a relationship found between the returned device and the reported incident. A visual inspection was performed on the exterior of product and no physical damage was observed. A burnt resistor was found on the power supply. The light source failed power up. Cause of power failure is a defective power supply/ballast. Unit powered up and passed functional testing with a known good power supply installed. The complaint was confirmed and the investigation has concluded the cause of the failure to be a defective electronic component. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported that the light source had a burnt resistor, resulting in overheating. No patient injury reported. Procedure completed with the same device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameRFB LIGHT SOURCE 500XL XENON
Type of DeviceIMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7044280
MDR Text Key139552898
Report Number1643264-2017-01740
Device Sequence Number1
Product Code FFS
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,other,use
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/01/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200568F
Device Catalogue Number72200568F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-