Model Number SC-2352-70 |
Device Problems
Fracture (1260); Migration or Expulsion of Device (1395); Material Protrusion/Extrusion (2979)
|
Patient Problems
Inadequate Pain Relief (2388); No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/14/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model: sc-2352-70 serial: (b)(4), description: linear 3-4 lead, 70cm.
|
|
Event Description
|
A report was received that the patient underwent a lead revision procedure due to lead migration of both leads.During the revision, the physician noticed that one of the leads had a rough edge over the end electrodes.Upon closer examination, it appeared that some of the wires were protruding.The lead was replaced and both leads were repositioned.The patient was stable post-operatively.
|
|
Manufacturer Narrative
|
Additional information was received that the patient experienced inadequate therapy due to lead migration.Analysis of the explanted lead sc-2352-70 (b)(4) revealed that the complaint has been confirmed.Visual inspection of the lead found that the lead proximal end is fractured between contacts 3 and 4.There are no exposed cables at the fractured portion of the proximal end.The cause of the proximal array damage could not be determined.
|
|
Event Description
|
A report was received that the patient underwent a lead revision procedure due to lead migration of both leads.During the revision, the physician noticed that one of the leads had a rough edge over the end electrodes.Upon closer examination, it appeared that some of the wires were protruding.The lead was replaced and both leads were repositioned.The patient was stable post-operatively.
|
|
Event Description
|
A report was received that the patient underwent a lead revision procedure due to lead migration of both leads.During the revision, the physician noticed that one of the leads had a rough edge over the end electrodes.Upon closer examination, it appeared that some of the wires were protruding.The lead was replaced and both leads were repositioned.The patient was stable post-operatively.
|
|
Event Description
|
A report was received that the patient underwent a lead revision procedure due to lead migration of both leads.During the revision, the physician noticed that one of the leads had a rough edge over the end electrodes.Upon closer examination, it appeared that some of the wires were protruding.The lead was replaced and both leads were repositioned.The patient was stable post-operatively.
|
|
Search Alerts/Recalls
|