No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that a monitor cable was replaced to resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect cable has been received by the manufacturer for evaluation.The received cable was found to have no apparent damage and passed continuity test with no opens.No fault found.
|
A medtronic representative reported that, while outside of a procedure, the display on the navigation system monitor "flickered".Once the flickering sensation passed, the monitor would intermittently lose display functionality.There was no patient present when this issue was identified.No additional information was provided.
|