A received image of the reported co2 absorber bypass valve confirms the reported issue.The absorber bypass valves are located in the patient cassette where the co2 absorber is docked, one at the inlet and one at the outlet.The main task of these valves is to lead the re-breathed patient gas through the co2 absorber in order to remove the co2 from the gas.The absorber bypass valves are equipped with spring-loaded cut-off valves that will automatically close the absorber bypass valves when the co2 absorber is bypassed and removed.With the absorber bypass valves closed, the bypass channel inside the cassette will be open.During the cleaning procedure of the patient cassette, these valves are removed from the patient cassette and put on a tool to be cleaned.To put the absorber bypass valves on the tool, the spring needs to be stretched.If the spring is stretched too much when inserting the tool, the spring may me deformed.This may lead to the valve not being able to close properly when the co2 absorber is in bypassed position.This fault will not affect the system function while the co2 absorber is docked in place.This fault can visually be noticed during assembling of the patient cassette after cleaning.Our conclusion is that the spring in the absorber bypass valve was stretched too much by the user during the cleaning procedure.
|