ST. JUDE MEDICAL, INC. CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM2100
Device Problems Communication or Transmission Problem (2896); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/04/2017
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined but is consistent with the change in device position.
 
Event Description
During follow-up, the device was unable to be paired with the patient activator.It was suspected that the device had rotated from its original implant position.No intervention were performed and the patient was stable and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7061437
MDR Text Key93081105
Report Number2017865-2017-34368
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K122161
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Model NumberDM2100
Device Catalogue NumberDM2100
Device Lot NumberS000008734
Other Device ID Number05414734502597
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-