Model Number PM3262 |
Device Problem
Pacemaker Found in Back-Up Mode (1440)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/16/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the pulse generator was in back-up mode pre-implant.About a month prior, a firmware upgrade was performed but was not in back-up at that time.The device was not used.
|
|
Manufacturer Narrative
|
The reported field event of backup vvi (bvvi) was confirmed in the laboratory and was due to a failed product code download.The device was tested on the bench and after reloading product code indicated normal device characteristics.
|
|