Model Number PM3262
Device Problem Pacemaker Found in Back-Up Mode (1440)
Patient Problem No Patient Involvement (2645)
Event Date 10/16/2017
Event Type  malfunction  
Manufacturer Narrative
The reported field event of backup vvi (bvvi) was confirmed in the laboratory and was due to a failed product code download. The device was tested on the bench and after reloading product code indicated normal device characteristics.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the pulse generator was in back-up mode pre-implant. About a month prior, a firmware upgrade was performed but was not in back-up at that time. The device was not used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUADRA ALLURE MP RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7064702
MDR Text Key112589778
Report Number2017865-2017-35125
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/08/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/31/2019
Device Model NumberPM3262
Device Lot NumberA000042846
Other Device ID Number05414734509091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-