• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Sticking (1597)
Patient Problem No Patient Involvement (2645)
Event Date 11/03/2017
Event Type  malfunction  
Manufacturer Narrative
A gehc service representative performed a checkout of the equipment and confirmed the reported complaint.Inspection of the equipment noted that the tabletop was interfering with the function of the oxygen flush valve.The tabletop was adjusted.No report of patient involvement.(b)(4).
 
Event Description
The hospital reported that, during the preoperative checkout, it was noted that the oxygen flush button would stick open, there was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
540 w. northwest highway
barrington, IL 
MDR Report Key7067902
MDR Text Key93435006
Report Number2112667-2017-02251
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-