Model Number IS4000
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2017
Event Type  malfunction  
Manufacturer Narrative
No part(s) will be returned to intuitive surgical, inc.(isi) for failure analysis investigations.An isi field service engineer was dispatched to the facility and was able to reproduce the reported failure.The fse reprogrammed the video processor which resolved the reported issue.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
 
Event Description
It was reported that during a da vinci-assisted prostatectomy procedure, the customer experienced multiple error messages when they connected the endoscope.Intuitive surgical, inc.(isi) technical support assisted the customer with troubleshooting, however, the issue persisted.The patient had already been administered anesthesia and the surgical ports had been placed when the issue occurred.At that time, the surgeon made the decision to abort the procedure.There was no report of patient harm, adverse outcome or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI XI SURGICAL SYSTEM
Type of Device
ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key7071524
MDR Text Key93735690
Report Number2955842-2017-00808
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIS4000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-