No part(s) will be returned to intuitive surgical, inc.(isi) for failure analysis investigations.An isi field service engineer was dispatched to the facility and was able to reproduce the reported failure.The fse reprogrammed the video processor which resolved the reported issue.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable after the start of a surgical procedure.Although no patient harm occurred, if this malfunction were to recur it could cause or contribute to an adverse event.
|
It was reported that during a da vinci-assisted prostatectomy procedure, the customer experienced multiple error messages when they connected the endoscope.Intuitive surgical, inc.(isi) technical support assisted the customer with troubleshooting, however, the issue persisted.The patient had already been administered anesthesia and the surgical ports had been placed when the issue occurred.At that time, the surgeon made the decision to abort the procedure.There was no report of patient harm, adverse outcome or injury.
|