Model Number UNKNOWN-S
Device Problem Inaccurate Flow Rate (1249)
Patient Problem Hematoma (1884)
Event Type  Injury  
Event Description
It was reported that the physician has had several patients with overdrainage of cerebrospinal fluid (csf) even when the valve was set to a performance level of 2. 5. It was noted there has been a 70% increase in incident lately due to overshunting along with an increase in patients presenting with subdural hematomas during the last 2-4 months. The physician has had to "tie off" a few shunts recently too because the 2. 5 setting was not slow enough. The physician has not had any instances where they have explanted the device from these affected patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN STRATA VALVE/SHUNT
Type of DeviceSHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key7073025
MDR Text Key93418926
Report Number2021898-2017-00618
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial
Report Date 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberUNKNOWN-S
Device Catalogue NumberUNKNOWN-S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/06/2017
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

Patient Treatment Data
Date Received: 11/30/2017 Patient Sequence Number: 1
-
-