The product was disposed of by the hospital and is no longer available for return.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a medical procedure using a neuron max 6f 088 long sheath (neuron max).During the procedure, the physician advanced the neuron max through a non-penumbra 8f sheath without any resistance.Next, the physician attempted to advance a neuron select 6f select catheter (6f select) through the neuron max but encountered resistance and decided to retract the 6f select.While retracting the 6f select, the physician encountered resistance but was able to remove the 6f select.The physician then replaced the neuron max with a new one and used it with the same 6f select and non-penumbra sheath to complete the procedure.There was no report of an adverse effect to the patient.
|