Model Number FUSION
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/08/2017
Event Type  malfunction  
Manufacturer Narrative
The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided. The reported issue is suspected to have been caused by the excessive volume of patient exams on the navigation system.
 
Manufacturer Narrative
Patient information not provided due to (b)(6) patient privacy regulations. A medtronic representative inspected the navigation system onsite and deleted the patients from the past multiple years and the system is now functioning well. No parts were replaced or returned to the manufacturer for evaluation.
 
Event Description
A site representative reported that, while in a functional endoscopic sinus surgery (fess) procedure, the navigation system had difficulty tracking the instruments. It took multiple times to get registered and in navigate, the system was having difficulty picking up the instruments and seeing them to change instruments. The site confirmed that all equipment was away from the emitter and had reduced metal as much as possible. There was a reported delay to the procedure of 45 minutes due to this issue and no impact on the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7085199
MDR Text Key102832331
Report Number1723170-2017-04966
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/05/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/23/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-