• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-1
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/08/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, the receiver was overheating.No additional event or patient information is available.Data was received but further investigation cannot be performed to confirm the problem and determine the root cause of the reported issue.No product was provided for evaluation.The complaint confirmation of the overheating of the receiver could not be determined.A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The receiver was returned for evaluation.A visual exterior inspection was performed and passed.The receiver was charged and will boot.The receiver log was reviewed, finding no errors related to the complaint.A receiver functional test was performed and passed all relevant testing.The receiver was opened for an internal inspection and passed.The complaint of a receiver ceasing to function could not be confirmed.The root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key7089040
MDR Text Key94295894
Report Number3004753838-2017-110044
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT20649-1
Device Catalogue NumberSTK-GL-010
Device Lot Number5225150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age55 YR
Patient Weight57
-
-