Model Number N/A
Device Problem Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed. Visual examination of the returned parts state signs of repeated use and missing ball bearings. Device history record was reviewed and no discrepancies were found. The device was subsequently re-designed to account for this failure mode with a new locking mechanism. It was determined that this device was manufactured prior to this design change. The root cause of the reported issue is attributed to design deficiency. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
(b)(4). The complaint device has been returned, but the device investigation has not yet been completed. Once the evaluation is completed, a supplemental medwatch 3500a will be submitted. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the provisional was missing ths spring and bearing. No adverse events have been reported as a result of the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePERSONA CONSTRAINED TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM
Type of DevicePROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7089695
MDR Text Key195834822
Report Number0001822565-2017-08351
Device Sequence Number1
Product Code OIY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK123459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/15/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number42527900404
Device Lot Number62836622
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown
Removal/Correction NumberZ-1052-2015

-
-