PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problems Computer Software Problem (1112); Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2017
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that bluetooth connection was lost during the case. The screen was black with an error notification after login on the tablet, and refresh was needed after restart. The black screen again appeared later, with the need to login. During the lead analysis application, the hourglass stayed, and the application was restarted. Because the lead analysis was needed when the application was down, the legacy analyzer was used to complete the lead analysis. No patient complications have been reported as a result of this event. The mobile programmer application remains in use.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARELINK SMARTSYNC
Type of DeviceANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7090916
MDR Text Key94646073
Report Number3004593495-2017-00678
Device Sequence Number1
Product Code DTC
Combination Product (Y/N)N
Reporter Country CodeNL
PMA/PMN NumberK163008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial
Report Date 12/06/2017
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/07/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number24970A
Device Catalogue Number24970A
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/19/2017
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-