• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Component Missing (2306); Failure to Calibrate (2440)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/27/2017
Event Type  malfunction  
Manufacturer Narrative
Product event summary: at analysis the customer comment that the stylus would not calibrate was confirmed but the comment that the disk was stuck was not confirmed.It was noted that "disk eject" is difficult on these programmers but that it is normal for this model.It was additionally noted that no electrocardiogram (ecg) leads were returned with the programmer.The connection from the xy board to the overlay was reseated and the stylus calibrated, ecg cable and leads were added, the hard drive reconfigured and the software was reloaded and updated.The programmer then passed its final functional and system tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer stylus would not calibrate and a disk was stuck.The programmer was also missing electrocardiogram (ecg) cables.The programmer was returned for service.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7093739
MDR Text Key94644060
Report Number2182208-2017-02086
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994581570
UDI-Public00613994581570
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received10/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/23/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-