During a battery replacement surgery, a patient's replacement device was found to have low impedance.The surgeon tried troubleshooting by reinserting the lead pin into the connector block multiple times and then performing an impedance test using a test resistor.The results of the impedance test with the test resistor displayed normal impedance, indicating no generator malfunction.The surgeon indicated that there was no noticeable lead damage or break, or suspected reason for low impedance.The battery was still implanted, but the lead was not replaced.No surgical intervention has occurred to date.No additional relevant information has been received to date.
|