Model Number S7
Device Problems Computer Software Problem (1112); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/10/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. The representative reported that the image would load on the newer version of the application software. It was noted that the application software was upgraded on the navigation system. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional.
 
Event Description
A medtronic representative reported that, while loading exams for a procedure, the orientation of the exams was displaying differently in the application software than their true identity. Axial images were showing as sagittal, coronal were showing as axial, and sagittal were showing as coronal. Burning new cds did not resolve the reported issue. Electronic picture archiving and communication systems (pacs) confirmed the orientation on their system. Restarting the navigation system did not resolve the reported issue. Once the images were loaded via unstructured digital intercommunication of medicine (dicom) functionality was restored. There was a reported delay to the procedure of more than 1 hour due to this issue. There was no impact on patient outcome. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S7 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7102144
MDR Text Key94516321
Report Number1723170-2017-05038
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial
Report Date 12/08/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-